Needle damage
C A U T I O N
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
Before a syringe is inserted, the position Change Syr should first be verified. The
position Change Syringe should not be located above an object where the needle
could collide with a vial, for example.
Path:
Menu > Utilitites > SyringeF3Change Pos.
Move the Injection Unit to a location where no collision can occur with the Z-axis by
selecting the X- and Y- axes appropriately. The value for the Z-axis is given as a
default and a change of this position is not necessary in standard operation.
This precaution helps avoid needle damage during routine operation. Nevertheless,
the description below on how to install the DLW Syringe Assembly recommends
switching off the unit and inserting the assembly without using the command
Change Syringe.
LC Injector HTC/HTS Description and Installation
3
Installation