LC Injector HTC/HTS Description and Installation
Special Functions
Item
Offset Z
Path Offset X
Path Offset Y
Path Offset Z
dxRow
dYRow
dzRow
dxCol
142
Description
The Z offset is defined as the distance from the plate surface of the Tray
Holder to the teaching point, black needle guide flush underneath plate of
Tray Holder (which corresponds to plate thickness of Tray Holder).
Measured in mm.
This item is available at User Level in section Utilities/Tray and can be used
for fine tuning a Tray installed in a Tray Holder. Important for well-plates 384
installed in a Stack.
N O T E
Position#1 is the only correction point.
The Path Offset X can be specific for a Tray, adjustment in X-direction.
Expressed in mm.
N O T E
This item is only valid if a Path is assigned to the Tray Holder as
specified in this combination.
The Path Offset Y can be specific for a Tray, adjustment in Y-direction.
Expressed in mm.
N O T E
This item is only valid if a Path is assigned to the Tray Holder as
specified in this combination.
The Path Offset Z can be specific for a Tray, adjustment in Z-direction.
Expressed in mm.
N O T E
This item is only valid if a Path is assigned to the Tray Holder as
specified in this combination.
1
Correction for inclination of a Tray in X-row direction. For details, see
"Correction for Sample Position"
1
Correction for inclination of a Tray in Y-row direction. For details, see
"Correction for Sample Position"
1
Correction for inclination of a Tray in Z-row direction. For details, see
"Correction for Sample Position"
1
Correction for inclination of a Tray in X-column direction. For details, see
"Correction for Sample Position"
on page 94.
on page 94.
on page 94.
on page 94.
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual