Table 33
Parameter
Syringe
Sample Volume
Airgap Volume
Front Volume
Rear Volume
Needle Gap Valve Clean
Fill Speed
Fill Strokes
Stator Wash
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
Repeatability Test with UV-Detection
Value
100 µL DLW
7 µL/10 µL
3 µL
5 µL
5 µL
3 mm
5 µL/s
0
0
LC Injector HTC/HTS Description and Installation
Parameter
Pullup Del
Inject to
Inject Speed
Pre Inj Del
Pst Inj Del
Valve Clean Time Solvent 2
Post Clean Time Solvent 2
Valve Clean Time Solvent 1
Post Clean Time solvent 1
3
Troubleshooting
Value
500 ms
Specified Inj.
Valve
5 µL/s
500 ms
500 ms
2 s
2 s
2 s
2 s
165