Agilent Technologies 1290 Infinity User Manual page 192

Lc injectors htc/hts
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

4
Appendices
DLW Principles and Operating Details
Table 36
Macro Description
The valve is rinsed with Wash Solvent 2,
followed by Wash Solvent 1
The Injection Valve is switched back to position
Standby.
End of macro DLW Priming.
192
DLW Priming
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
Macro Variable
Stator Wash Time Solvent 2
Stator Wash Time Solvent 1
N O T E
The loop is filled with the last rinse with
Wash Solvent 1. Verify the composition of
Wash Solvent1. The solvent should have a
lower elution power than the solvent
gradient starting conditions or sample
solvent composition. This is important for
partial loop filling.