LC Injector HTC/HTS Description and Installation
Object Positions
A possible inclination of the Tray column can be corrected in the same manner
as described above.
Figure 47
The three corner points are now adjusted for a possible inclination of the Tray
in any axis. A possible deviation from an ideal axis position for the other vials,
caused by variance from the horizontal, is interpolated by the system without
further teaching.
For routine work using standard vials it is not necessary to compensate for a possible
N O T E
inclination as long as the vial detection is within the Z-Tolerance range.
96
Adjusting Column Inclination at Third Tray Corner Position
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual