LC Injector HTC/HTS Description and Installation
Object Positions
To check the next corner of the Tray, activate function key F3 again, example
Tray Type VT32, position 008.
The next information provided by the pop-up window asks to correct a
possible inclination of the Row. Continuing the dialog provides the user the
opportunity to adjust the three axes perfectly to match the vial top – or in case
of a well-plate, the top of the plate.
Figure 46
Adjusting Row Inclination at second Tray corner position
The last step will be to move to the next corner of the Tray by again activating
function key F3. This will be position 32 in the example of Tray Type VT32.