LC Injector HTC/HTS Description and Installation
Installation
Mounting the Pump Holder
The pump holder always has to be mounted on the far left of the X-axis. The
solvent lines are connected to the holder. Moving across the pump holder with
the Z-axis could kink the stabilizing wire. To install the holder, dismounting
the drive from the LC Injector HTC/HTS X-axis is recommended.
1 Locate the pump holder and the two supplied Torx screws.
2 Attach the holder to the X-axis as shown in
66
Figure 19
Attaching Pump Holder Bracket to X-Axis
Figure 19
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
on page 66.