If the basic LC Injector HTC/HTS System is installed, than a single power
supply is installed only. Turning OFF the power supply or pulling this single
power cord in an emergency case will stop the complete LC Injector HTC/HTS
System.
It is important that the power supply (ies) are in a location where the power
ON and OFF switch is accessible and easy to operate, and where it is possible
to unplug the AC power cord from the power supply/wall outlet in case of
emergency.
Other Hazards
To avoid injury during LC Injector HTC/HTS System operation, keep your
hands away from the syringe.
Do not operate the LC Injector HTC/HTS System without the safety shield. The
safety shield must be installed for safe operation.
To avoid injury, observe safe laboratory practice when you handle solvents,
change tubing, or operate the LC Injector HTC/HTS System. Know the physical
and chemical properties of the solvents you use. See the Material Safety Sheets
from the manufacturer for the solvents in use.
Use caution when working with any polymer tubing under pressure:
• Always wear eye protection when near pressurized polymer tubing.
• Do not use polymer tubing that has been severely stressed or kinked.
• Do not use polymer tubing, in particular not PEEK or Tefzel tubing, with
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
Tetrahydrofuran (THF), Dimethylsulfoxid (DMSO), chlorinated organic
solvents, concentrated mineral acids, such as Nitric, Phosphoric or Sulfuric
acids, or any related compounds to above listings.
General Information
Safety Information
1
9