LC Injector HTC/HTS Description and Installation
Object Positions
Waste Reference Point
The reference position for the Waste Port – a hole (slightly larger than the
needle guide) – has been placed in the middle position of Wash1 and Wash2.
(See
hole with the bottom of the lower needle guide lightly touching the surface of
the Wash Station assembly.
Figure 50
The Waste position represents an Injector within the software. It is defined in the Object
N O T E
class Injectors.
1 On the handheld controller select the sequence:
2 Enter Waste and adjust the X-, Y-, Z-positions until the lower needle guide of
3 Press F1 to verify the position and F4 for Home.
100
Figure 50
on page 100.) The lower needle guide should be centered in this
Waste Reference Position
Menu/Setup/Objects/Injectors/Waste 1 .
the Injection Unit centers on top of the Wash Station assembly. Fine tune
the Z-Position until the bottom of the lower needle guide is flush with the
top surface.
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual