Agilent Technologies 1290 Infinity User Manual page 139

Lc injectors htc/hts
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

Item
Spl1 Offset X
Spl1 Offset Y
Spl1 Offset Z
1
2
The dimensions used for Row/Column Length or X-, Y-, Z-Offsets for Sample Number 1, are
N O T E
specific for a particular Tray Type. If a general adjustment for a Tray from a specific vendor –
e.g. for a Deepwell Plate – has to be made, this should be done in the Tray Type class of
Object. For fine tuning, the Offset items from the class Trays can be used.
Explanation for various Patterns in Tray Type:
1 Pattern Type Regular.
2 Pattern Type Staggrd+.
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
Description
The X offset is defined as the distance from the outer edge of the Tray in
the X-direction to the center of the vial position 1. Measured in mm.
The Y offset is defined as the distance from the outer edge of the Tray in
Y-direction to the center of the vial position 1. Measured in mm.
The Z offset can be selected for special cases where, e.g. the syringe
needle has to reach a deeper point before penetration into a vial. A Z offset
value = 0 considers the Tray surface or vial top as the reference point.
Measured in mm.
Available beginning with Firmware Level 4.1.X
Available beginning with Firmware Level 2.5.X
The sample or well positions are equal distributed.
Figure 68
Pattern Tray Type Regular
The sample or well positions are arranged in a staggered pattern. Offset of
second Row is shifted by +50 % of hole pattern.
LC Injector HTC/HTS Description and Installation
3
Special Functions
139