3
LC Injector HTC/HTS Description and Installation
Troubleshooting
Unexplained collisions of the Injection Unit.
Probable cause
1
Object Not Detected (Object e.g. Tray1, LCVlv1, Stk1- 01)
Probable cause
1
2
Object Collision before Z Tolerance (Object e.g. Tray1, LCVlv1,
Stk1-01)
Probable cause
1
150
Object(s) not defined correctly.
Incorrect Z value for the affected Object (e.g.
LCVlv1, THldr1 or CStk1-01).
Missing sample vial, Deepwell- or
Microtiterplate.
Incorrect Z value for the affected Object.
Suggested actions
Define Object(s) correctly. (See
Reference Positions"
Suggested actions
Redefine Z value for the affected Object. (See
"Defining Object Reference Positions"
page 87).
Check position and insert, if necessary, vial,
deepwell- or microtiter plate.
Suggested actions
Redefine Z value for the affected object. (See
"Defining Object Reference Positions"
page 87).
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
"Defining Object
on page 87).
on
on