Pilot Distance Incorrect
When making a new tube connection it is absolutely vital to follow the following guidelines.
N O T E
Failure to observe this procedure will lead to an incorrect pilot distance, which will result in
N O T E
dead volume – actual dead volume and not delay volume.
1 Slip the nut and ferrule over the tube. The end of the tube should stick out
2 Press the tube end firmly into the female counterpart for the connection.
3 Move the nut and ferrule together into the female counterpart for
4 Tighten the nut as described above.
Figure 72
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
approximately 50 mm (2 inch).
connection. At this stage is it important to maintain pressure on the tube to
avoid the tube slipping backwards, out of position.
Pressing Tube into Place
LC Injector HTC/HTS Description and Installation
3
Troubleshooting
155