Agilent Technologies 1290 Infinity User Manual page 94

Lc injectors htc/hts
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

3
LC Injector HTC/HTS Description and Installation
Object Positions
Correction for Sample Position
The following description of the correction is done by using the path to the
Utilities section.
After selecting the Utilities function Trays, verify whether the Tray Type
assigned corresponds to the Tray placed into the Tray Holder. If this is
verified, the vial positions can be checked at the three corners of the Tray.
This is activated by the F3 function key.
A dialog window opens, displaying a note to Adjust offset X-, Y-, Z for Position 001.
Continuing in this dialog will provide the user the opportunity to adjust all
three axes perfectly by the top of the vial. Please note that this step is only
valid for position number 001.
Figure 45
94
Adjusting Offset X,Y,Z for Tray Position 001
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual