Service
Service items may be changed by selecting the particular item (see
page 131).
Item
Check Motors
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
LC Injector HTC/HTS Description and Installation
Description
Every stepper motor can be moved separately. This function is useful to
check out a specific motor or to perform an endurance test after a repair.
The Start and Stop positions of the test path can be selected. The motor
Current, Speed, and Acceleration are selectable (accept default value to be
sure). A Pause time can be specified. A Cycle Limit and Actual Cycle
Counter are useful tools for endurance tests.
Phase A: This item is used to verify the functionality of the stepper motor.
The current can be manually increased for tests. Used for troubleshooting
only.
Phase B: This item is used to verify the functionality of the stepper motor.
The current can be manually increased for tests. Used for troubleshooting
only.
MicroStep Inx: The motor can be moved within a range of up to four full
steps by applying the assigned current.
N O T E
These items are typically used for service. If applied, do not leave
the test settings for longer than a few minutes. If current setting is
in a critical range the motor can be damaged.