Agilent Technologies 1290 Infinity User Manual page 81

Lc injectors htc/hts
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

Injection Valve Plumbing and Connection to DLW Wash Station
1 Connect the injection valve Waste Line from port 2 (example standard valve
If two injection valves are used in the LC Injector HTC/HTS configuration, the second
N O T E
waste line can be connected to the lower front waste port of the DLW Wash Station. If only
one connector is used, apply a dummy plug to the second port on the Wash Station.
2 Place the Waste Tube from the DLW Wash Station into the waste container.
The waste container must be positioned below the injection valve. Make sure that the
N O T E
waste liquid flows into the waste container without restriction.
Figure 32
The Agilent specific Rheodyne Valve with Agilent PNo. 5067-412 is connected in the same
N O T E
way to the HPLC system as shown in
valve description.
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
Cheminert type), as shown in
of the Wash Station
Solvent Line Plumbing on Rheodyne Valve PD Series
LC Injector HTC/HTS Description and Installation
Figure 32
on page 81, to the front Waste Port
"Installation"
on page 108. For details see the specific
3
Installation
81