Specifications
Table 27
Valve Stator:
Rotor:
Bore Size:
Volume for Vertical Port:
Ports and Engraving:
Valve Volume Full Loop:
Connections:
Maximum Pressure:
Standard Loop:
RoHS:
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
LC Injector HTC/HTS Description and Installation
Specifications
SPC-5 coated stainless stator
RPC-13 Polyimide based rotor seal
Vertical Port: 0.25 mm (0.010 in)
Ports 2 to 6: 0.20 mm (0.008 in)
0.75 µL
Ports 2 to 6: 0.07 µL
Engraving: 0.05 µL (standard engraving)
Full Loop: 0.87 µL
Partial Loop: 0.82 µL
10-32 male threaded fittings, 1/16 in tubing OD
1200 bar (18000 psi; 120 MPa)
2 µL; Loop ID: 0.13 mm (0.005 in)
Compliant
3