LC Injector HTC/HTS Description and Installation
Troubleshooting
Re-use of Installed Nuts/Ferrules
Do not re-use an installed nut/ferrule for any other connection. Any tube with installed
N O T E
nut/ferrule used for one purpose should not be installed for any other application.
1 Cut a portion of the tube end until straight tubing is reached and remove
This statement is valid for a loop as well. Never re-use a loop for another valve or a valve
N O T E
after exchanging the stator.
If a finger-tight nut has been installed, replace it as well.
N O T E
If a stainless steel nut has been used, ensure that the correct type is used, see
Nuts or Ferrules used"
Newly installed finger-tight ferrules can hold a backpressure of up to 200 bar.
N O T E
154
the ferrule.
on page 158.
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
"Wrong