LC Injector HTC/HTS Description and Installation
Special Functions
Item
Test Head
Test Switches
132
Description
This tests the functionality of the Injection Unit (Head). Two tests can be
activated by the function keys:
F1Check Plg: The stepper motor of the plunger drive is tested. Sensor
check.
F2Check Z: The vial detection sensor and Needle Guide Blocking function
is tested.
These tests are only required if any erratic behavior of the System is
observed. In such cases contact a representative of Agilent Technologies.
X-Limit = Actual status of X-axis sensor. Move the X-Y-Carriage manually
across the limit switch to verify activation of the sensor.
Y-Limit = Actual status of X-axis sensor. Move the Y-axis manually to the
Y-zero point (at X-axis), crossing over the limit switch to verify activation of
the sensor.
Z-Limit = Actual status of Z-axis sensor. Move the syringe slider assembly
manually up to the zero Z-axis point, crossing over the limit switch to verify
activation of the sensor.
Aux1 Limit = Actual status of Aux1 limit sensor
Aux2 Limit = Actual status of Aux2 limit sensor
Plg –Pos = Actual status of Plg –Pos
Needle Guide = Actual status of needle guide sensor
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual