HDD Initialization Procedure
1. Items to be prepared
1) PC
Be sure that the version of the Service Support Tool that supports the host machine is
installed.
2) Crossover Ethernet cable
2. Preparing for the Installation of the System Software of Host machine
1) If the PC and host machine have been started, turn the main power OFF.
2) Connect the PC and the machine using an Cross Ethernet cable.
3) Turn ON the power of the PC.
3. Registering the system software
1) Insert the latest System Software CD into the PC using the SST.
2) Start the SST.
3) Click 'Register Firmware'.
4) Select the drive where the system software CD has been inserted, and click the [SEARCH]
button.
5) Click the [REGISTER] button.
6) Click [OK].
4. Initializing HDD
< In case of SST >
1) Start the host machine with download mode in safe mode.
2) Start the SST.
3) Select the model. Then, select [Single] and click [Start].
4) Execute [Format HDD].
5) After formatting is completed, shut down and restart the host machine.
6) Terminate the SST.
<In case of USB flash drive>
1) Connect the USB memory to the PC.
2) Start up SST, and click the USB icon displayed in the target selection screen.
3) Select the drive, the model series, and the version to be written to the USB flash drive, and
click [Confirm].
4) Click [Start], and remove the USB flash drive when the version has been written to the USB
flash drive.
5) Terminate the SST.
9
Installation > HDD Data Encryption Kit-C6 > Checking the Security Mark
6) Connect the USB memory device to the host machine, and start the host machine with
download mode in safe mode.
7) Press keys on the Control Panel in the order shown below.
• [4] : Clear/Format
• [2] : Flash Format
• [0] : OK
• [Reset] : Start shutdown sequence
• [C] : Return to main menu
• [Reset] : Start shutdown sequence
• [0] : OK (The power of the host machine is turned OFF automatically.)
8) Turn ON the power of the host machine.
9) After the host machine has restarted, remove the USB flash drive from the mount mark (
) on the bottom right of the screen.
Checking the Security Version
1) Press the Counter Check key on the control panel.
2) Press the [Check Device Configuration] key appearing on the control panel.
3) Confirm that '2.00' or '2.01' is displayed for the [Canon MFP Security Chip] item indicating
the version of the security chip.
If multiple Encryption Boards are installed, the version information for each board is
displayed.
CAUTION:
The user can refer to the version displayed in the [Canon MFP Security Chip] item
indicating the version of the security chip to confirm that an Encryption Board that
contains a security chip version that has received CC certification is operating correctly.
Checking the Security Mark
The user can check the security mark displayed on the Control Panel when using the host
machine to confirm that security is being protected. This security mark is displayed only when
an Encryption Board is installed and operating normally. The location where the security mark
is displayed is described in the User's Guide as shown below.
< Confirming the Security Mark >
When the HDD Data Encryption Kit is operating normally, a security mark (
on the lower left corner of a panel screen.
9
Installation > HDD Data Encryption Kit-C6 > Checking the Security Mark
9-55
) is displayed
9-55