Interlocking Function - Nikon Eclipse Ni-E Instruction Manual

Upright motorized microscope (focusing nosepiece system)
Hide thumbs Also See for Eclipse Ni-E:
Table of Contents

Advertisement

Motorized Unit that Save/Loads the State with MODE Function
Setting items
CONDEN.
Motorized universal condenser's condenser module currently in the optical path (not supported for
focusing nosepiece) (The motorized condenser is not used in the focusing nosepiece system.)
FL TURRET
Filter cube of the motorized epi-fluorescence cube turret in the optical path
FL [SHUTTER]
Open/closed state of the motorized epi-fluorescence cube turret's built-in shutter
FL 2nd
Filter cube of the motorized epi-fluorescence cube turret on the 2nd layer that is brought in the
optical path
FL2 [SHUTTER]
Opening/closing of the built-in shutter of the motorized epi-fluorescence cube turret on the 2nd layer
EX WHEEL
Motorized excitation filter wheel's excitation filter currently in the optical path
BA WHEEL
Motorized barrier filter wheel's barrier filter currently in the optical path
PATH
Optical path in the motorized quadrocular tilting tube
Z
Position of the elevating section
EPI SHUTTER
Opening/closing of the EPI motorized shutter
DIA SHUTTER
Opening/closing of the DIA motorized shutter
LAMP ADJ.
Brightness of the dia-illumination lamp
LAMP
Turns the dia-illumination lamp on or off.
INTSL [SHUTTER]
Opening/closing of the motorized HG precentered fiber illuminator's built-in shutter
INTSL
ND filters in the motorized HG precentered fiber illuminator
A. STOP
Aperture diaphragm diameter of the motorized universal condenser (not supported for focusing
nosepiece) (The motorized condenser is not used in the focusing nosepiece system.)
F. STOP
Field diaphragm diameter
ZOOM
Zoom magnification of the motorized DSC zoom port
ND WHEEL
ND filter transmittance of the motorized ND filter wheel
*1
The elevating section of the Ni-E is automatically excluded from the target of mode setting upon recycling the power
on the microscope main body, so as to avoid contact between the sample and the objective.

Interlocking Function

18.2
With the Ni-E focusing nosepiece system, the rotation of the motorized quadrocular tilting tube and the motorized
epi-fluorescence cube turret can be set to interlock when pressing the [CAPTURE] button to capture an image.
For interlocking settings, see "20 Operation on DS-L3" - "20.1 Setting Up the Microscope - (4) Configuring the Movement"
or Chapter 6 "5 Configuring the Movement of Motorized Devices ([MOVEMENT] Screen)" in the "Microscope Control"
manual bundled with DS-L3.
Setting items
PATH
(Optical path of the motorized
quadrocular tilting tube)
FL TURRET
FL 2nd
(cube inside motorized
epi-fluorescence cube turret)
Interlocking in Sleep mode
No interlocking takes place in Sleep mode.
Chapter 3
Individual Operations
Stored Information
*1
Interlocking
• FRONT (tube adapter)
• REAR (rear port)
• OFF (Do not interlock as factory
setting)
• Bring the specified filter cube into the
optical path.
• OFF (Do not interlock as factory
setting)
82
Interlock condition
Motorized quadrocular tilting tube must be
connected
Motorized epi-fluorescence cube turret must
be connected