IR-DIC Microscopy
5
Fully open the field diaphragm and aperture diaphragm.
Press the left side of the DIA field diaphragm button (
mark side) and fully open the field diaphragm.
Pressing the right side (
Turn the aperture diaphragm lever fully clockwise to open
the aperture diaphragm completely.
6
Set the condenser turret to [O] position (empty: bright-field).
Turn the condenser turret and face the display [O] on the
turret to the front. The turret's empty position enters the
optical path in this state.
Chapter 2
Microscopy Operations
Preparation
Procedure
1
2
3
4
5
6
7
8
mark) will close the diaphragm.
15
Focus and Orientation
Microscopy
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 ■
DIA field diaphragm button
Field diaphragm fully open
- - _ _ - - _ _ _ _
O P E N
_ _ _ _ _ _ _ _ _ _ _ _
Aperture
diaphragm
lever
Aperture diaphragm fully open
[F. STOP] button, sub screen: DIA field diaphragm
Display [O]
Moving the condenser to the empty position
ND
ND
NCB
8
32
11
OUT
IN
Z :
_ _ _ _ _
0 . 0 0 0 u m
F S 3 0 . 6 m m
open