Nikon Eclipse Ni-E Instruction Manual page 102

Upright motorized microscope (focusing nosepiece system)
Hide thumbs Also See for Eclipse Ni-E:
Table of Contents

Advertisement

4
Shown below is the actual settings. See (1) through (7).
Saving the setting
After completing the setting, be sure to press the
[SAVE] button on the [MICROSCOPE SETUP
MENU] screen to save the settings. If the DS-L3 is
turned off without saving settings, previous values
are restored. See Chapter 6 "1 Bulk Saving of
Settings" in the DS-L3 instruction manual
"Microscope Operation" for the details of saving.
(1) Setting Optical Elements Installation Information (See DS-L3 Chapter 6 "2 Configuring Optical
Elements Information".)
Configure optical elements used for the following motorized units:
<Default settings>
Address 1 of the motorized universal condenser and the motorized barrier filter wheel is set to [OPEN] by default and
cannot be changed. The rest of all has not yet been set. [OPEN] setting is available only for Address 1 in a
motorized/intelligent epi-fluorescence cube turret.
Device icons on the top of the screen are as shown below from
left.
• Motorized nosepiece/intelligent nosepiece: Objective settings
(not used in the focusing nosepiece system)
• Motorized condenser: DIC module, PH module settings (not
used in the focusing nosepiece system)
• Motorized/intelligent epi-fluorescence cube turret (on the 1st
layer): Filter cube settings
• Motorized/intelligent epi-fluorescence cube turret (on the 2nd
layer): Filter cube settings
• Motorized excitation filter wheel: Excitation filter settings
• Motorized barrier filter wheel: Barrier filter settings
Press a device icon button, then press the address button of an
optical element to select the optical element from a list of
registered elements.
If you cannot find an optical element you want to set in the list,
pressing the [OPTIONAL] button allows you to add up to 10
optical element information. Follow the instruction on the screen
during the operation. Use the same procedure to set it again
when an optical element attachment has changed.
You can use the same procedure for any devices to set the
optical element information.
Chapter 3
Individual Operations
88
Saving the setting
[SETUP MENU] → [COMPONENTS]
Device selection
Example: Epi-fluorescence cube turret, address 4
setting
Selecting a filter cube