Using Operation Buttons On Ni-E; Front Operation Button - Nikon Eclipse Ni-E Instruction Manual

Upright motorized microscope (focusing nosepiece system)
Hide thumbs Also See for Eclipse Ni-E:
Table of Contents

Advertisement

2

Using Operation Buttons on Ni-E

Various operation buttons are provided on the front, right, and left of the Ni-E main body. The following describes the
function of each button (factory default setting):

Front Operation Button

2.1
7
6
No.
Name
1
Display panel
2
DISPLAY Previous
button
3
DISPLAY Next button
4
Brightness Control
button
5
Z-RESET button
6
to
11
FUNCTION button
12
BINO optical path
button
13
FRONT optical path
button
14
REAR optical path
button
15
Power LED
16
Sleep LED
17
DISPLAY switch button
Chapter 3
1
8
4
2
17
12
15
Front Operation Button
Displays either the "microscope state" or "FUNCTION button's function". Use the
"DISPLAY SWITCH" button to select the display.
Switch the microscope state on the display panel to the previous pattern.
Switch the microscope state on the display panel to the next pattern.
Change the brightness of the display panel. The brightness changes in the following
order by pressing the button: Bright → Dark → Off → Bright → ...
Reset the Z-axis coordinate value on the display panel to 0.000 μm.
*
Button 1, 2: Rotate the condenser CCW, CW. (The motorized condenser is not used
in the focusing nosepiece system. Change the function of the button to a different
operation.)
Button 3, 4: Rotate the excitation filter wheel CCW, CW.
Button 5, 6: Rotate the barrier filter wheel CCW, CW.
Direct 100% of the optical to the binocular (with status indicator LED, binocular 100%
when LED ON, vertical tube or rear port 100% when LED OFF).
Direct 100% of the optical to the vertical tube (with status indicator LED, vertical tube
100% when LED ON, binocular or rear port 100% when LED OFF).
Direct 100% of the optical to the rear port (with status indicator LED, rear port 100%
when LED ON, binocular or vertical tube 100% when LED OFF).
Displays the power ON/OFF status of the main body and control box A. The LED
lights up when the power is on.
The LED is on when the device is in sleep mode (where the power supply to
motorized devices is suspended to minimize the generation of noise). Turn on/off the
sleep mode from DS-L3.
Switch the "microscope state" and "FUNCTION button's function" display.
Individual Operations
13
14
16
Function
45
9
10
11
5
3