Exit Calibration Mode; Restore Factory Calibration - Xantrex GPIB-M-HPD Operating Manual

Multichannel functionality for programmable dc power supplies
Table of Contents

Advertisement

Exit calibration mode

When you have completed calibration, exit calibration mode by sending the
command:
CAL:STAT OFF, "0000"
where "0000" is replaced with your security code.

Restore Factory Calibration

Restore the unit to the calibration constants set at the factory by sending the
command:
CALibration:RESTore
Note This procedure should not be used in place of regular calibration, but may be useful to
restore the unit to an operational state in case of failure.
Release 2.1
Calibration
Exit calibration mode
129