Invacare
- TDX-SP / TDX-SP N
You will be able to see the strip fuses (2) once the fuse holder (1) is open.
If the strip fuse has blown, you must first ascertain and rectify the cause of the fault.
The main fuse may only be replaced once the fault has been rectified.
Remove the strip fuse nuts (3) with an 8 mm socket spanner.
Replace the strip fuse.
Fitting the main fuse:
Refit all the parts again in reverse order.
To conclude, check all the mobility aid functions.
8.2.6

Replacing the main fuse (TDX-SP N)

Note
To replace the fuse, the battery cable with the fuse attached to it must be replaced.
Replace the battery cables, as described in chapter 8.2.3.