SERVICE MANUAL
●
Rotate the top walking beam (5) so that you can insert the screw (4) through the drillhole (3)
on the gas cylinder into the gas cylinder and walking beam.
●
Push the bolt (4) through. Fix the bolt with a 5/16" Allen key.
●
Fit the nut (6) with a 1/4" socket spanner onto the bolt and tighten the nut to 8.5 Nm.
●
Fit both the anti-tip system protective caps.
●
Fit the drive wheel as described in Chapter 8.13.1.
●
Replace the batteries in their correct location as described in Chapters 8.2.2/8.2.3 and 8.2.1.
●
Remove the wooden block from under the mobility aid.
122
®
Invacare
- TDX-SP / TDX-SP N