Rotate the top walking beam (5) so that you can reach the screw (4) on the gas cylinder
through the drillhole (3) in the battery case.
●
Remove the nut (6) with a 1/4" socket spanner. Use a 5/16" Allen key to counter the bolt while
doing so.
●
Push the bolt (4) towards the battery case so that the gas cylinder (1) is separated from the
walking beam.
●
Push the gas cylinder against the battery case.
●
Raise the rear axle so that the gas cylinder is activated and then retract the cylinder.
●
Push the gas cylinder to the rear and pull it off the battery case.
SERVICE MANUAL
117