Invacare
- TDX-SP / TDX-SP N
ATTENTION: Danger of crushing!
The motor gearbox unit is extremely heavy. Injury hazards to hands!
● Please note the high weight
●
Pull the motor/transmission unit (4) off the bolt.
Refitting the motor/transmission unit:
●
Replace the defective parts.
●
Refit the motor/transmission unit in reverse order.
●
Tighten the motor bolts (5) to 30 Nm.
SERVICE MANUAL
133