SERVICE MANUAL
Make a marking on the motor and the carbon brushes in order to guarantee correct refitting.
Remove the carbon brushes complete from the mounting (2).
Check the carbon brushes and springs for the level of wear, broken components or
discolouration.
Refitting the carbon brushes:
Depending on the condition of the brushes and the spring:
- either replace the brushes in exactly the same position from which they were taken or
- fit new brushes.
Replace the plastic caps and tighten them.
Refit the top walking beams as described in Chapter 8.18.2.
Refit the bottom walking beams as described in Chapter 8.17.
Refit the drive wheels as described in Chapter 8.13.1.
Note
The following procedure is necessary to run the carbon brushes in after replacement and to
therefore guarantee maximum performance.
WARNING: Risk of accidents!
Hazards to workers, surroundings and mobility device!
● Do not leave the mobility device unattended during the following procedure!
● Stellen Sie sicher, dass BEIDE Antriebsräder angehoben werden und den Boden nicht
berühren können.
● Secure the area.
Lift the mobility aid up on one side and get a second person to place a wooden block around
14 cm high underneath it so that the drive wheel is suspended freely.
Repeat the same procedure on the other side of the mobility aid.
Allow the motors to run in the forward direction for an hour.
Allow the motors to cool down for 30 minutes.
Allow the motors to run in the reverse direction for an hour.
Lift the mobility aid off the wooden blocks.