8.16.5 Adjusting the setting screw

Requirements:
● 1/4" socket spanner
● M8 washer
Remove both the anti-tip system protective caps.
The protective caps on the anti-tip system will either be held using a screw or a cable tie
depending on the mobility aid's date of manufacture.
Ensure that the mobility aid is contacting the floor with all its wheels and castors.
Loosen and remove the two locknuts (2) with a 1/4" socket spanner.
Note
For the following operation, use an M8 washer as a thickness gauge.
Rotate the setting screw (3) until it has a space of 1.5 mm (A) between it and the gas cylinder
(4).
Tip: use an M8 washer as a thickness gauge.
Fit the lock nut (2) and tighten it using a 1/4" socket spanner.
Refit the two protective caps again.