SERVICE MANUAL
PROBLEM
Motors do
not run
Motor
makes
clicking
noise
Scraping
noise or
motor
blocked
Transmissi
on makes
clicking
noise
Transmissi
on losing
oil
26
OTHER
POSSIBLE
SYMPTOMS
CAUSE
None
Bad connections Check all cabling.
Fuse burnt out
Batteries
defective
Cabling to
electronic
module or
joystick
defective
Electronic
module
defective
Corroded
Water, salt or
contacts
urine has
penetrated
None
Clutch(es)
defective
Bearing
defective
Collector
defective
None
Clutch(es)
defective
Bearing
defective
Transmission
defective
None
Transmission
defective
Drive wheel
loose
None
Sealing ring on
drive shaft
defective
®
Invacare
- TDX-SP / TDX-SP N
SOLUTION
Documentation
See chapter 8.2.7.
Check cabling and
See chapters 8.2.7 and
replace fuse
8.2.5/8.2.6
Replace batteries
See chapter 8.2.2/8.2.3.
check cabling
See chapter 8.2.7.
Replacing the
See chapter 8.6.2.
electronic module
Check cabling,
See chapter 8.2.7.
replace if
necessary.
Replace clutch
See chapter
8.19.4/8.19.5.
Replacing the
See chapter
motor
8.19.1/8.19.2.
Measure internal
See chapters 7.1 and
resistance of
8.19.1/8.19.2
motor, replace
motor if defective.
Replace clutch
See chapter
8.19.4/8.19.5.
Replacing the
See chapters
motor
8.19.1/8.19.2 and
8.19.4/8.19.5
Replace
See chapters
transmission.
8.19.1/8.19.2 and
8.19.4/8.19.5
Replace
See chapters
transmission.
8.19.1/8.19.2 and
8.19.4/8.19.5
Tighten drive
See chapter
wheel, secure
8.19.1/8.19.2.
bolts with Loctite if
necessary
Replace
See chapters
transmission if
8.19.1/8.19.2 and
sealing ring
8.19.4/8.19.5
defective.
Check carbon
See chapters
brushes for oil
8.19.1/8.19.2,
wetting, replace
8.19.4/8.19.5 and
motor if brushes
8.19.6/8.19.7
wet