4
Invacare
- STORM
7.3.4

Checking and replacing the main fuse

CAUTION: Fire hazard!
 Always use an original strip fuse with the approved amperage.
 If the main fuse has blown, first rectify the cause before fitting a new one.
CAUTION: Fire hazard!
Fitting the incorrect strip fuse causes a fire hazard!
 Only fix the strip fuses in the sequence shown in the image on the right!
 Tighten the nuts with 3.3 or 3.5 Nm!
Requirements:
 8 mm socket spanner
Note
If the fuseholder is damaged, you can replace this complete with the battery cables.
Dismantling the main fuse:
Remove the battery compartment cover, disconnect the battery plugs and pull the batteries out
of the battery compartment as described in chapter 7.3.1.
The fuseholder (1) is located on top of the batterie.
Open the fuseholder cover.
MAX
SERVICE MANUAL
1. Strip fuse
2. Spade terminal
3. DIN 6923 nut
49