SERVICE MANUAL
●
Undo and remove the lock nut (2), the setting screw (4) with a 1/4" socket spanner.
●
Rotate the setting screw (4) towards the battery case until it can be removed from its holder
(3).
Refitting the setting screw:
●
Screw the new setting screw (4) into the holder (3).
●
Screw the two bolts (2) together with their nylon washers (3) in again.
●
Tighten the bolts using a 1/4" socket spanner.
●
Remove the wooden block from under the mobility aid so that all wheels and castors are in
contact with the floor.
124
®
Invacare
- TDX-SP / TDX-SP N