SERVICE MANUAL
6.1

Drive fault diagnosis

PROBLEM
Mobility aid
will not
start
Mobility aid
judders in
drive mode
Mobility aid
pulls to left
or right
Error
message
does not go
out
Motors stop
and start
irregularly
24
OTHER
POSSIBLE
SYMPTOMS
CAUSE
The remote
Drive motors
status display
disengaged
illuminates
normally and
displays blink
code 5/6.
Remote status
Batteries
display does not
defective
illuminate
Completely
discharge
battery
Power supply to
remote
interrupted
Remote
defective
Remote status
Various causes
display blinking
None
Batteries
defective
(unstable
voltage)
Drive motor(s)
defective
None
Drive motors
running
asymmetrically
None
Bad connections Check all
Motor brake
defective
None
Voltage decline
®
Invacare
- TDX-SP / TDX-SP N
SOLUTION
Documentation
Engage drive
See operating manual
motors
Replace batteries
See chapter 8.2.2/8.2.3.
Pre-charge
See operating manual
batteries
Check master fuse See chapter 8.2.5/8.2.6.
Check cables
See chapter 8.2.7.
between the
modules for loose
connections and
damage
Replace remote
See chapter 8.6.1.
Assess error code See chapter 6.2 and 6.3.
Replace batteries
See chapter 8.2.2/8.2.3.
Replace motor(s)
See chapter
8.19.1/8.19.2.
Replace carbon
See chapter
brushes
8.19.6/8.19.7.
Change
See chapter 8.7.
programming to
synchronise
motors.
See chapter 8.2.7.
connecting cables.
Measure internal
See chapters 7.2 and
resistance of
8.19.1/8.19.2
brakes, replace
motor if defective.
Stop journey and
allow electronic to
cool down.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents