SERVICE MANUAL
8.10.5 Replacing the foot plate on the centre-mounted legrest (electric & manual)
Requirements:
● 1/64" Allen key
● 1/4" Allen key
A manual legrest can be seen in the illustration. The method of fixation to the seat is almost
identical with both legrests.
Uninstalling the footplate
●
If possible, raise the lifter to the highest position.
●
If possible, tip the wheelchair with the tilt module as far backwards as it will go.
●
Switch the electronic control system off.
●
Use the 1/64" Allen key to loosen
and remove both screws (3).
●
Replace the foot plate (1).
Installing the footplate:
●
Assemble the parts in reverse order.
●
Tighten the hexagon socket head screws (3) till they are tight enough that the footplate can
barely be moved.
In the area denoted by the (2) there is an adjustment screw that can only be reached when the
footplate is swivelled upwards.
●
Use the 1/64" Allen key to adjust the angle of the footplate to the needs of the user with this
screw.
100
®
Invacare
- TDX-SP / TDX-SP N