Invacare TDX-SP Service Manual page 73

Invacare tdx-sp wheelchair service manual
Hide thumbs Also See for TDX-SP:
Table of Contents

Advertisement

®
Invacare
- TDX-SP / TDX-SP N
Refitting the seat carrier:
Hold the seat support (1) firmly while you fit the bolts (3) and the metal plates (2).
Screw the bolts in but do not tighten them.
Use the protractor to adjust the seat carrier to the required angle.
Tighten the bolts to 18 Nm.
Refit the batteries including battery carrier as described in Chapters 8.2.2/8.2.3 and 8.2.1.
Refit the seat.
Adjust the anti-tip system as described in Chapter 8.16.5.
SERVICE MANUAL
73

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents