Service Plan (1X Annually) - Invacare TDX-SP Service Manual

Invacare tdx-sp wheelchair service manual
Hide thumbs Also See for TDX-SP:
Table of Contents

Advertisement

SERVICE MANUAL
5

Service plan (1x annually)

CAUTION!
Danger of injury and damage to property, if the maximum speed reduction on a
wheelchair with a lifter does not function correctly!
The wheelchair's control unit must reduce the maximum possible speed as soon as the
lifter is raised.
● Test the maximum speed reduction for correct function after any maintenance work or
modifications to the wheelchair.
Component
Safety belt
Armrests
Side panels
Seat angle
adjustment
Power backrest
(if fitted)
Frames
(chassis) /
battery
mounting
Anti-tip system
20
Check
Damage to safety belt
Belt lock function
Damage to armrests
Armrest fixings
Damage to side panels
Side panel fixings
Tight seating of SL
fuses
Damage to backrest
Seams
Fixing
Check cable
Check function
Check fixings, welded
seams and battery
mounting
Check gas cylinder
Check adjustment
Invacare
Remedy
Replace belt if
damaged.
Replace belt if
damaged.
Replace covering if
damaged
Tighten screws
Replace side panels
if damaged
Tighten screws
Replace SL fuses if
necessary
Replace parts if
damaged
Tighten screws
Replace cable motor
if necessary
Tighten screws
Replace components
if necessary
Repair or replace if
damaged
Replace gas cylinder
every 2 years!
Speed
®
- TDX-SP / TDX-SP N
Notes
See chapter 8.5.
See chapter 8.5.
See chapter 8.16.
See chapter 8.16.5.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents