8.5

Checking and replacing the main fuse

CAUTION: Fire hazard!
· Always use an original strip fuse with the approved amperage.
· If the main fuse has blown, first rectify the cause before fitting a new one.
CAUTION: Fire hazard!
Fitting the incorrect strip fuse causes a fire hazard!
· Only fix the strip fuses in the sequence shown in the image on the right!
· Tighten the nuts with 3.3 or 3.5 Nm!
Requirements:
· 8 mm socket spanner
Note
If the fuseholder is damaged, you can replace this complete with the battery cables.
Dismantling the main fuse:
·
Remove the battery compartment cover, disconnect the battery plugs and pull the batteries out
of the battery compartment as described in chapter 8.4.1.
·
The fuseholder (1) is located on top of the batterie.
·
Open the fuseholder cover.