Remove the two knurled screws (1) including washers.
Remove the cover (2).
Pull the plug out of the actuator module.
Open the cable holder on the underside of the seat.
Uncover the actuator cable so that you can pull it out of the lifter module later on together with
the actuator.
WARNING: Danger of crushing!
The seat is very heavy. Injury hazard to hands if the seat is unintentionally folded down!
● You should seek help from a second person.
Ensure that the seat is properly secured against unintentional folding or falling down before
you start removing the actuator.
If you are unable to secure the seat, you must remove it completely.
If you are working on a combined lifter/tilt module, the lifter actuator is located in the lower section
of the module.
Remove the retaining ring (1) with the retaining ring pliers.
Push the bolts (4) out of the guides. When removing the front bolts, take care with the bushes
(3).
Remove the actuator from the lifter module.
Refitting the lifter actuator:
Refit all parts in the reverse sequence
Carry out a performance test.
SERVICE MANUAL