SERVICE MANUAL
Open the lifter energy chain:
- lift the cover (1) up.
- pull the tab (2) out of the cover.
- repeat this part by part.
Remove the two cable ties which fix the cable to the ends of the energy chain.
Pull the cable out of the energy chain.
Undo the screws (1) with a crosshead screwdriver.
Remove the energy chain (2).
Undo the screws (4) with a crosshead screwdriver.
Remove the magnets (3).
Remove the sensor (5).
Undo the four bolts (2) on both sides of the seat using a ½" socket spanner.
Lift the lifter/tilt module off the battery case.
Refitting the lifter/tilt module:
Refit the lifter/tilt module in the reverse sequence.
Tighten the connecting bolts between the lifter/tilt module and the battery case to 18 Nm.