7.8

Checking and replacing the main fuse

CAUTION: Fire hazard!
A short circuit can cause extremely high currents which can result in spark formation and
fire!
· Always use an original strip fuse with the approved amperage.
· If the main fuse has blown, first rectify the cause before fitting a new one.
CAUTION: Fire hazard!
Fitting the incorrect strip fuse causes a fire hazard!
· Only fix the strip fuses in the sequence shown in the image on the right!
· Tighten the nuts with 3.3 to 3.5 Nm!
Requirements:
· Phillips screwdriver
· Strip fuse
· socket spanner, 8 mm
· torque wrench 0-20 Nm (or similar)
Note
A spare fuse is located behind the Invacare logo on the rear battery.
Note
If the fuseholder is damaged, you can replace this complete with the battery cables.
· Remove both battery cases as described in
Chapter 7.7.1.
· Remove the covers on both battery cases as
described in Chapter 7.7.2.
68
®
Invacare
- Bora/Spectra XTR
1. Strip fuse
2. Spade terminal
3. DIN 6923 nut