Invacare TDX-SP Service Manual page 25

Invacare tdx-sp wheelchair service manual
Hide thumbs Also See for TDX-SP:
Table of Contents

Advertisement

®
Invacare
- TDX-SP / TDX-SP N
PROBLEM
Motor runs
but loses
power
Motors stop
and do not
start again
Motors lose
power
during
journey
Motor
judders or
runs
irregularly,
or only one
motor runs
OTHER
POSSIBLE
SYMPTOMS
CAUSE
None
High motor
loading causes
electronic
module to
decrease
voltage
None
High motor
loading causes
electronic
module to
decrease
voltage
None
Fuse burnt out
None
Motor defective
None
Electronic
module
defective
None
Bad connections Switch mobility aid
None
Carbon brushes
worn
Clutch(es)
defective
Bearing
defective
Collector
defective
SOLUTION
Documentation
Stop journey and
allow electronic to
cool down.
Leave mobility aid
switched on and
allow electronic
module to work.
Charge batteries
overnight with
mobility aid
switched on.
Check cabling and
See chapters 8.2.7 and
replace fuse
8.2.5/8.2.6
Check carbon
See chapter
brushes and
8.19.6/8.19.7.
replace if
necessary
Measure internal
See chapters 7.1 and
resistance of
8.19.1/8.19.2
motor, replace
motor if defective.
Replacing the
See chapter 8.6.2.
electronic module
See chapter 8.2.7.
off, wait 10
seconds, switch
mobility aid on
again. Check all
cabling.
Check carbon
See chapter
brushes and
8.19.6/8.19.7.
replace if
necessary
Replace clutch
See chapter
8.19.4/8.19.5.
Replacing the
See chapter
motor
8.19.1/8.19.2.
Measure internal
See chapters 7.1 and
resistance of
8.19.1/8.19.2
motor, replace
motor if defective.
SERVICE MANUAL
25

Hide quick links:

Advertisement

Table of Contents
loading