Removing the batteries:
Make the batteries accessible as described in Chapter 8.2.1.
Remove the terminal caps.
Remove the terminal bolts (3) with an 11 mm socket spanner. First remove the negative
terminal bolt. After this, you can remove the positive terminal bolt.
Remove the connection cable (4) and the connection cable with the main fuse (4).
Open the battery strap (5).
Lift the batteries (6) out of the battery carrier (1).
Refitting the batteries:
Refit the batteries in reverse order.
Ensure that the battery cage sockets/plugs have been correctly refitted. A polarity diagram is
located in the battery case cover.