Replacing The Motor/Transmission Unit (Ssd Motor) - Invacare TDX-SP Service Manual

Invacare tdx-sp wheelchair service manual
Hide thumbs Also See for TDX-SP:
Table of Contents

Advertisement

SERVICE MANUAL

8.19.2 Replacing the motor/transmission unit (SSD motor)

WARNING: Danger of crushing!
The wheelchair is very heavy. Danger of injury to the hands and feet!
● Let a second person help you.
Danger of injury due to uncontrolled movement of the wheelchair!
● Switch the vehicle's power system off (ON-/OFF key).
● Engage the drive motors.
● Secure the vehicle against rolling away by placing wedges under the wheels.
ATTENTION: Danger of crushing!
The motor gearbox unit is extremely heavy. Injury hazards to hands!
● Please note the high weight
Requirements:
● 6 mm Allenkey
● 5/8" Allen key
● 7/8" Allen key
● 5/16" Allen key
● Phillips screwdriver size 2
● 13 mm socket spanner
● 1/2" socket spanner
● 1/4" socket spanner
● 5/16" socket spanner
● torque wrench 5 - 25 Nm (or similar)
● torque wrench 10 - 80 Nm (or similar)
● two long wooden blocks, min. 14 x 14 x 30 cm
Uninstalling the motor / gear box unit:
Remove the rear battery box cover, as described in chapter 8.1.2.
Disconnect the motor plug from the power module.
Open the cable ties so you can freely access the motor cable.
Remove the drive wheel, as described in chapter 8.13.1.
Remove the wheel hub, as described in chapter 8.13.4.
134
®
Invacare
- TDX-SP / TDX-SP N

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents