Operating steps
1. As shown in the above figure, connect the load to the DUT. Ensure that the
polarity of the connections is correct.
2. Press Shift and CP key to enter the system utility function interface. Press
"Config" and "Sense" to enable (disable) the Sense function.
Note: The Sense terminals must be connected to the output terminals of
the DUT. If not, then the load cannot accurately measure the voltage of
the terminals in any function.
Tip: In case where the DUT outputs large currents, the load leads
should be as short as possible. Twisting the sense leads can also
assist in minimizing measurement noise.

External Trigger Function

The SDL features an external trigger which allows sequencing and control to
be performed using another instrument or trigger source. To start, set the
trigger source to "External" by pressing Shift > Utiity > Config > Page 2/2 and
set Trig to Ext. The trigger signal input and ground connection are shown on
the rear panel. The trigger input signal should be 0-5V TTL level
62