7.4.1 Current Setup Menu
Details of the current leak detector setup are displayed. Reject and audio set point status,
auto sequencer status & split flow. The calibrated leak setting, leak rate range settings and
vent lock out status.
a
Condition
Audio Set Point
See Section 7.4.1 "Current Setup
Menu" on page 128.
Auto Sequencer
See Section 7.4.1 "Current Setup
Menu" on page 128.
Calibration Set-Up
See Section 4 "Calibration Screen"
on page 65.
Leak Rate Range
See Section 4 "Calibration Screen"
on page 65.
128
Agilent Technologies Helium Leak Detector
Figure 7-15 Current Setup Screen
Table 7-3 Current Setup Screen Conditions
Display
Audio Set Point:
Active
Audio Set Point:
Inactive
Auto Sequencer: On
Auto Sequencer: Off
Int Cal Leak
Ext Cal Leak
Auto Ranging
Manual Ranging
Description
Audio Set Point is active.
Audio Set Point is inactive.
Auto Sequencer (automatic test cycle)
is enabled.
Auto Sequencer (automatic test cycle)
is disabled.
System is set to calibrate to the internal
standard leak.
System is set up to calibrate to an
external standard leak.
System is set to auto range through
the leak rate scale.
System is set for manual range control.