Additional Error Reduction by Connecting a Series Resistor

The additional error due to the DC voltage source noise can be reduced by connecting a series
resistor. Figure 8-6 to Figure 8-9 show esminations of 10% additional error when the 0 (no
series resistor), 5 k , 10 k , 50 k ,
and 100 k series resistor is connected.
DC voltage source performance
Ripple
Wide band noise
Switching noise
Figure 8-6. 10 % Additional Error (Measurement Time: 10 msec)
8-12
General Information
5  Vrms(@50 Hz)
p
1  Vrms/
Hz (@50 Hz)
10 mVrms(@100 Hz)