Power Meter Operation
Making Measurements using Sensor Calibration Tables
tables are currently selected. The "Sensor Tbls" screen is displayed as
shown in Figure 2-4.

Making the Measurement

To make the power measurement, you zero and calibrate the power meter,
then set it for the frequency of the signal you want to measure. The power
meter automatically selects the calibration factor from the sensor
calibration table.
Note
2-22
Figure 2-4: "Sensor Tbls" Screen
The following procedure details the calibration for channel A. To
calibrate channel B use the same procedure using the equivalent
channel B softkeys. To calibrate both channels, zero the power
meter using
Zero Both
channel A then repeat step "3" through to step "9" using the
equivalent channel B softkeys.
Zero
1. Press
.
Cal
2. Press
. During zeroing, which takes approximately 10
Zero A
seconds, the wait symbol is displayed.
3. Press
. The setting of the reference calibration factor is
Cal
obtained from the sensor calibration table and displayed under the
softkey.
A Ref CF
4. Connect the power sensor to the POWER REF output.
5. Press
to calibrate the power meter. During calibration the
Cal A
wait symbol is displayed. (The power meter automatically turns
on the POWER REF output.)
in step "2", then follow this procedure for
Agilent E4419B User's Guide