Calibrate At The End Of A Long Cable - Agilent Technologies FieldFox N9912A User Manual

Rf analyzer
Hide thumbs Also See for FieldFox N9912A:
Table of Contents

Advertisement

Verifying Phase Accuracy
 Connect a LOAD standard at the calibration reference plane (where calibration
 In NA Mode, select a S11 Reflection with Polar or Smith Chart format.
 Because all LOAD standards have delay, you should see a small amount of
 In general, the measurement result should agree with the characteristics of the
Test the Jumper Cable
With the LOAD standard still connected, move the jumper cable while observing
the trace.
 If the measurement trace is relatively stable, the jumper cable is of good
 If you observe significant movement in the peaks of the measurement trace
Using QuickCal to Calibrate at the End of a Long Cable
The QuickCal process requires that there be sufficient resolution (data points) to
unwrap the cable reflection phase correctly. This requirement is met by having a
step frequency (frequency span between data points) with a wavelength = (4*
length of cable(m)) / (velocity factor).
For a 5 meter cable with a velocity factor of 0.7:
wavelength = 4* 5m / 0.7
wavelength = 28.57
step freq = speed of light/wavelength.
step freq = 3e8 / 28.5
step freq = 10.5 MHz
Resolution = frequency span / step freq
For a 2 MHz to 6 GHz sweep:
Resolution = 5.998e9 / 10.5e6
Resolution = 571 points or greater.
On the FieldFox, select 601 points before calibrating.
In addition, for VERY long cables (100 ft) there may be too much loss to obtain
distinct measurements of the open, short and load.
56
Lower than 0 dB indicates a GOOD calibration.
o
Higher than 0 dB indicates that you should recalibrate.
o
standards were connected).
phase rotation as a function of frequency.
calibration standard.
quality.
when moving the cable (>5 dB), the jumper cable may need to be replaced.
N9912A FieldFox User's Guide

Advertisement

Table of Contents
loading

Table of Contents