Click Enter Curve/Linearize (fig. 4.25). The Densitometer
Data Entry window appears (fig. 4.27).
Fig. 4.27 Densitometer Data Entry Window
2.
Measure individual patches on the calibration page by placing the
page on a white background; then, center the densitometer's lens
over the patch and either click the black button on top of the
XD-1 or click the measure button on the Densitometer Data Entry
window (fig. 4.26). The measurement appears in the box to the
right of the Measure button.
3.
Make sure to enter the solid black (1) and solid white (2)
measurements first. Enter the number of the patch whose
densitometer measurement most closely matches (within
2%) the corresponding percentage in the enter the box numbers
section (3).
For example, when you are reading for 2%, if the box labeled 7
measures 2%, enter the number 7 in the corresponding box.